Elastomeric Closures Treatment

Thanks to a wide experience in the pharmaceutical industry, Fedegari has developed a special sterilizer addressed to elastomeric closures treatment. This process equipment can perform washing, rinsing, siliconizing, sterilizing, drying and automatic unloading of stoppers in a sterile area. All these processes are managed by Thema4 process controller.

The stopper treatment autoclave uses a drum with a special magnetic coupling, which guarantees more space in the chamber.

As for Fedegari washer sterilizer (FOWS) this is a dual-purpose unit: when is not running as a stopper treatment it can be used as a normal moist-heat sterilizer.

  • AISI 316L stainless steel chamber.
  • Chamber range from 200 to 6.800 liters, but loading capacity is fully customizable, according to customer’s need.
  • Chamber internal surface is mechanically polished and complies with rigorous pharmaceutical standards, with a roughness Ra <0,4 μm.
  • 6 PT100 RTD temperature sensors to monitor chamber and product temperature (3 flexible, 2 fixed + 1 located in the door-chamber jacket).
  • Two doors with hinged opening systems to cope with available space in the smartest possible way (unloading side comes with semi-automatic door when chute is installed).